Arium® Advance RO
Distributor & Service Provider In India

(Reverse Osmosis System With iJust)

The Arium® Advance RO provides high-quality Type 3 reverse osmosis water. By using the integrated, innovative iJust function, it does not only optimize water consumption but also ensures that natural resources are used conscientiously. In contrast to conventional water purification systems, the unique touch display with intuitive menu navigation makes it extremely easy to use.

With a flow rate of up to 8, 16 or 24 L/h, the automatic RO membrane backflush and a constant flow rate, the Arium® Advance RO is the optimal choice for general laboratory applications.

Advantages

  • Time saving
  • Optimized water consumption
  • Easy to use

Applications

  • Buffer and media preparation
  • Water for Laboratory devices (Autoclaves)

Technical specifications

Dimensions: Width × Height × Depth 35.0 × 50.1 × 45.1 cm
Empty weight Approx. 15 kg
Operating weight Approx. 22 kg
Power supply 100–240 VAC (±10%); 50–60 Hz, 130 VA (max.)
Operating temperature 2 °C–35 °C at max. 80% relative humidity
Storage temperature 5 °C–45 °C at max. 80% relative humidity

Product Water Quality

Water purification method Particle filtration, adsorption by means of spherical activated carbon, catalyst and reverse osmosis
Water type Type 3, reverse osmosis water
Production output 8, 16, or 24 L/h
Water dispensing flow rate Up to 3 L/min
Typical conductivity < 20 µS/cm
Typical resistivity < 20 µS/cm
Retention of dissolved organic substances (MW > 300 Dalton) > 99%
Particle and microorganism retention > 99%
Particle content No particles > 0.22 µm
Bacteria < 0.001 CFU/mL
Endotoxins < 0.001 EU/mL
RNase concentration < 1 pg/mL
DNase concentration < 5 pg/mL
Water dispensing flow rate Up to 2 L/min
Volume-controlled dispensing 0.05–50 L in 50 mL steps

Feed water quality

Exclusively potable tap water pursuant to the drinking water standards of the USA, the European Union, or Japan.

Input Pressure 0.5–6.9 bar, Recommended > 2 bar
Temperature 2–30 °C
Specific Conductivity < 1,500 µS/cm (compensated to 25 °C)
TOC (Total Organic Carbon) < 2,000 ppb
Max. Total Hardness 360 ppm (max. CaCO3)
Free Chlorine < 4 ppm
Iron (Total Fe Content) < 0.1 ppm
Fouling Index (SDI) < 5
Turbidity < 1 NTU
pH Value 4–10