x
I n k a r p

Reichert® 2SPR

Reichert® 2SPR
Distributor & Service Provider In India

The Best 2-Channel SPR Value You Can Buy.

Choose the 2-channel surface plasmon resonance (SPR) system that lets you do more — and costs up to 30% less than the leading competitor. The Reichert2SPR System pushes the limits of detection and sensitivity, which is why it’s used by leading universities, academic facilities and biotech startups around the world for label-free biomolecular interaction analysis.

With Reichert, you get more, including:

  • High-quality, publication-ready data
  • The most sensitive, flexible and affordable 2-channel system available
  • The best performance-to-cost ratio for individual labs
  • A proven SPR leader since 1997

To see how Reichert can help advance your research and lead to new breakthroughs, contact us today to schedule a demo in your lab.

    Technical Data

Transducer Principle Kretschmann prism, multiple angles from fixed light input
Measurement Channels – Two (either parallel or series fluid connection)
Sample Loading Autosampler or semi-automatic injector, standard HLPC tubing and connectors, external syringe pump
Sample Capacity Any combination of up to 2 trays can be used. Choose: 12 (10 mL) or 48 (2 mL) vials; or 96-well (high or low) or 384-well plates
Sample Storage 4°C or ambient temperature
Flow Cell Volume (per channel) – 0.18 µL Flow Cell Surface Area (per channel) – 4.5 mm² (reference value)
Aspect Ratio > 25 (width/height)
Fluid Contact Materials Teflon™, Acetal Copolymer, PEEK™, Kalrez™, ETFE (Tefzal™)
Sample Volume 20-4500 µL (depends on installed loop volume)
System Fluid Volume 28 µL (typically) (0.01" I.D. tubing) or 7.5 µL (0.005" I.D. tubing)
Temperature Range 10°C below ambient to 70°C

  • MEASUREMENT SENSITIVITY
Refractive Index Resolution < 10-7 RIU (< 0.1 µRIU)
Refractive Index Range 1.33 to 1.40 (@780nm)
Analyte Concentration Range 1 mM to 1 pM
Minimum Molecular Weight Detection < 100 Daltons
Baseline Noise 0.1 µRIU peak-to-peak, 0.05 µRIU RMS, @ 25 mL/min
Baseline Drift < 0.1 µRIU/min

  • TYPICAL KINETIC AND EQUILIBRIUM CONSTANT RANGES
Association Rate Constant 103to 107M-1s-1
Dissociation Rate Constant 10-1to 10-5s-1
Equilibrium Dissociation Constant 0.1 mM to 1 pM
Electrical AC Power Supply, standard international voltage range w/universal adapter from 100 to 240 V; 50 to 60 Hz
Regulatory Compliance with the applicable sections of the European EMC Directive and IEC safety requirements for laboratory electrical equipment for measurement and control
Product Safety Compliance with IEC 61010-1 (Low Voltage Directive) under a Category classification EMC and Safety: CE mark certification (Class A, Type II)

  • Compare the Reichert4SPR to Cytiva Biacore™ T200

The Reichert2SPR - Reichert's SR7500DC system is the smart choice for scientists and lab managers. Reichert surface plasmon resonance instruments provide a highly sensitive and flexible platform to meet the demands of a variety of applications. Reichert’s low maintenance costs, autosampler capacity, and sensitivity make it the obvious choice.

Reichert2SPR Cytiva BiacoreTMX100
Number of Channels 2 2
DIY Fluidics Yes No
Autosampler Capacity Maximum 768 samples 15 samples
Baseline Drift 0.1 RU/min < 0.3 RU/min
Baseline Noise 0.05 RU(RMS) < 0.1 RU(RMS)
Specialized Flow Cells Electrochemistry, Photochemistry, Fluorescence, Mass Spectrometry None
Sample Compartment Temperature 4°C or Ambient Ambient
Flow Rate Range 0.1 –3000 µL/min 1 – 100 µL/min

Webinars

Upcoming Webinars

Master Polar Analyte Separation: Expert Tips for Chromatographers