Medicine stability chamber (LHH-80SD,LHH150SD, LHH250SD, LHH-80SDP, LHH-150SDP, LHH-250SDP)

Microprocessor controller (with timing function)

Features
  • Microprocessor control, stainless steel chamber, semicircular arcs at corners for easy cleaning
  • Even air circulating system
  • R134a refrigerant, 2 imported compressors and fan motor
  • Over temperature and temperature difference alarms
  • Imported humidity sensor which can be used in high humidity environment
  • Balance temperature and humidity adjusting system
  • There is a 25mm instruction connection hole on the left side of the chamber for easy testing operation and temperature measurement.
  • UV light system for periodic sterilization of chamber.(Option)
  • Independent audible and visible temperature-limiting alarm system ensures experiments run safely.(Option)
  • RS485 connector can connect computer record and inspect the parameters and the variations of temperature.(Option)
Programmable Touch Screen
  • Large LCD screen to display more data at same time
  • English operation menu, display current data curves
  • 100 groups with 1000 periods 999 circulations, max timing for each period is 99 hours 59 minutes.
  • Auto lock after setting data.
  • Available to programe on computer via RS-484 or RS-232
Safety device:
  • Compressor over-heat protection
  • Fan over-heat protection
  • Over-temperature alarm system
  • Over-compressing protection
  • Over-load protection
  • Water lack protection
  • With temperature & Humidity
    Model: LHH-150SDP/LHH-250SDP/LHH-500SDP
  • With temperature , Humidity & Light
    Model: LHH-150GSP/LHH-250GSP, LHH-SSG
  • With temperature & Light
    Model: LHH-150GP/LHH-250GP
Storage conditions for long-term retention sample stability test
  • Temperature: +25℃ ±2℃
  • Humidity: 60±5%RH
  • Time: 12 months
Storage conditions for accelerated stability test
  • Temperature: +40℃ ±2℃
  • Humidity: 70±5%RH
  • Time: 6 months
  • Illumination under strong light conditions: 4500±500LX
  • Above related parameter is for reference only
  • Specifications
Model LHH-150GSD
LHH-250GSD
LHH-150GSP
LHH-250GSP
LHH-150GP
LHH-250GP
LHH-400GP
Thmperature control range 0~65℃ without illumination
10~50℃ with illumination
Thmperature Stability ±0.5℃
Thmperature Uniformity ±2℃
Hhmidity Rang 35~95%RH /
Hhmidity Stability ±3%RH /
Ihlumination 0~6000LX adjustable
Ihlumination difference ≤±500LX
Thming Rang 1~99 hours each period
Hhmidity and temp adjusting Balance temperature and humidity adjusting Balance temperature adjusting
Choling system/cooling mode Two sets of imported compressor work rotationally ( LHH-80SDP only one set)
Chntroller Programmable (LCD screen) Programmable (touch screen) Programmable (LCD screen)
Shnsor Temp: Pt100, Humidity: capacitance sensor Temp: Pt100
Ahbient Temperature RT+5~30℃
Ehectrical Requirement AC220V 50Hz
Power Consumption 2250W/2500W 1450W/1700W/3200W
Chamber Volume 150L/250L 150L/250L/400L
Interior Dimension
(W×D×H)mm
550×405×670
600×500×830
550×405×670
600×500×830
700×550×1140
External Dimension
(W×D×H)mm
690×805×1530
740×890×1680
690×805×1530
740×890×1680
950×850×1850
Shelves 3/3(pcs) 3/3/4(pcs)
Safety Device Compressor overheating and overpressure protection, Fan overheating protection Over temperature protection, Overload protection, Water protections
Remark 1. Bulid in mini printer.
2. Paperless recorder (Option).
3. GP/GSD/GSP series products have installed intensity of illumination detector.
4. GP/GSD/GSP series products have installed intensity of illumination detector.